VersaPulse Laser - Small Power - Repair.

VersaPulse Laser

Another case of repair of the holmium laser VersaPulse Select II, brought from America.

Reason for repair - power Ho: YAG laser of a quickly weakened. The laser beam was of low power, but the device did not give any errors.

Power modes of the laser generator were normal. By slightly adding the laser operating voltage, the laser power increased slightly.

After disassembling the Ho: YAG generator, it turned out that its water channels are plugged with a salt stopper, this is because of the salt water. The laser quickly became unusable from overheating.

Ho: YAG laser generator with reflector removed.

VersaPulse Ho:YAG

Water Tube Fitting VersaPulse Laser.

VersaPulse Fitting

Ho: YAG laser generator - water channels.

VersaPulse water channels