VersaPulse Laser - Small Power - Repair.
Another case of repair of the holmium laser VersaPulse Select II, brought from America.
Reason for repair - power Ho: YAG laser of a quickly weakened. The laser beam was of low power, but the device did not give any errors.
Power modes of the laser generator were normal. By slightly adding the laser operating voltage, the laser power increased slightly.
After disassembling the Ho: YAG generator, it turned out that its water channels are plugged with a salt stopper, this is because of the salt water. The laser quickly became unusable from overheating.
Ho: YAG laser generator with reflector removed.
Water Tube Fitting VersaPulse Laser.
Ho: YAG laser generator - water channels.
Repair of laser surgical devices:
Surgical laser Lancet-1 scheme and structure.
Surgical laser Lancet-1 - Error 4 Repair.
|